IEC 60601 Medical electrical equipment

ID 11296 | | Visite: 15923 | Documenti Riservati Marcatura CEPermalink: https://www.certifico.com/id/11296

IEC 60601 x x

IEC 60601-1-X Medical electrical equipment

ID 11296 | State 08.2020

Stato e Preview allegati delle norme tecniche generali per apparecchi elettromedicali.

IEC 60601 è una serie di norme tecniche per la sicurezza e le prestazioni essenziali degli apparecchi elettromedicali, pubblicate dalla Commissione elettrotecnica internazionale.
Pubblicate per la prima volta nel 1977 e regolarmente aggiornate e ristrutturate, a partire dal 2011 sono attualmente costituite da uno standard generale IEC 60601-1, 8 standard collaterali (IEC 60601-1-X) e 80 standard particolari/performance ca (IEC 60601-2-X, IEC 60601-3-X).

IEC 60601-x-xx Example 

The IEC 60601-1-xx series of collateral standards for MEDICAL DEVICE ELECTRICAL EQUIPEMENT
The IEC 60601-2-xx series of particular standards for particular types of MEDICAL ELECTRICAL EQUIPMENT
The IEC 60601-3-xx series of performance standards for particular types of MEDICAL ELECTRICAL EQUIPEMNT

IEC 60601

La norma generale IEC 60601-1 - Apparecchi elettromedicali - Parte 1: Requisiti generali per la sicurezza di base e le prestazioni essenziali - stabilisce i requisiti generali della serie di norme. IEC 60601-1 è un punto di riferimento ampiamente accettato per le apparecchiature elettromedicali e la conformità a IEC 60601-1 è diventata un requisito per la commercializzazione di apparecchiature elettromedicali in molti paesi (in UE la norma è armonizzata per la Direttiva Dispositivi medici 93/42/CEE e il Regolamento Dispositivi medici (UE) 2017/745 che prevedono la marcatura CE).

La conformità a IEC 60601-1, è per le aziende di apparecchi elettromedicali, un requisito per la maggior parte dei mercati. Questo standard non garantisce l'efficacia di un dispositivo medico. Negli Stati Uniti, la FDA richiede prove di efficacia confermate tramite un'approvazione pre-mercato (PMA-
Premarket Approval)(1) o similare ad un dispositivo già autorizzato tramite una notifica pre-mercato 510 (k) (Premarket Notification 510(k)(2).

(1) Premarket approval (PMA) is the FDA process of scientific and regulatory review to evaluate the safety and effectiveness of Class III medical devices, and the most stringent of the device marketing applications. Class III devices are those that support or sustain human life, are of substantial importance in preventing impairment of human health, or which present a potential, unreasonable risk of illness or injury. Due to the level of risk associated with Class III devices, FDA has determined that general and special controls alone are insufficient to assure the safety and effectiveness of Class III devices. Therefore, these devices require a premarket approval (PMA) application under section 515 of the FD&C Act in order to obtain marketing clearance. Please note that some Class III preamendment devices may require a Class III 510(k). PMA applications will include technical sections, usually divided into non-clinical laboratory studies and clinical investigations. PMA approval typically requires a facility inspection to confirm compliance to 21 CFR 820 prior to approval.

(2) A 510(k) is the technical dossier required by the US Food and Drug Administration (FDA) to sell a medium-risk medical device or IVD in the United States. It is formally called a Pre-market Notification. A 510(k) contains detailed technical, safety, and performance information about a medical device. The documentation must demonstrate the device in question is "substantially equivalent" to a predicate device (i.e. a product already cleared for sale in the US). The FDA must review the 510(k) and "clear" your device before you can legally sell or distribute it in the United States.

510(k) submissions are required for most Class II (as well as some Class I and Class III) medical devices and IVDs. If you cannot determine the classification for your product, you can submit a 513(g), which is a written request to the FDA seeking classification confirmation.

Each person who wants to market in the U.S., a Class I, II, and III device intended for human use, for which a Premarket Approval (PMA) is not required, must submit a 510(k) submission to FDA unless the device is exempt from 510(k) requirements of the Federal Food, Drug, and Cosmetic Act (the Act) and does not exceed the limitations of exemptions in part .9 of the device classification regulation chapters (e.g., 21 CFR 862.9, 21 CFR 864.9). A 510(k) is a premarket submission made to FDA to demonstrate that the device to be marketed is at least as safe and effective (substantially equivalent) to a legally marketed device that is not subject to PMA. Submitters must compare their device to one or more similar legally marketed devices (predicates) and make and support their substantial equivalency claims. Until the submitter receives an order declaring a device SE, the submitter may not proceed to market the device. Once the device is determined to be SE, it can then be marketed in the U.S.

IEC Serie

IEC 60601-1:2005+AMD1:2012 (Armonizzzata MD)
IEC 60601-1-2:2014 (Armonizzzata MD)
IEC 60601-1-3:2008+AMD1:2013 (Armonizzzata MD)
IEC 60601-1-4:1996 (Ritirata)
IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 (Armonizzzata MD)
IEC 60601-1-8:2006+AMD1:2012+AMD2:2020  (Armonizzzata MD)
IEC 60601-1-9:2007+AMD1:2013+AMD2:2020 (Non Armonizzzata MD)
IEC 60601-1-10:2007+AMD1:2013+AMD2:2020  (Armonizzzata MD)
IEC 60601-1-11:2015+AMD1:2020  (Armonizzzata MD)
IEC 60601-1-12:2014+AMD1:2020  (Non Armonizzzata MD)
IEC TR 62348:2012

IEC 60601-1:2005+AMD1:2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1:2005+A1:2012 contains requirements concerning basic safety and essential performance that are generally applicable to medical electrical equipment. For certain types of medical electrical equipment, these requirements are either supplemented or modified by the special requirements of a collateral or particular standard. Where particular standards exist, this standard should not be used alone. This consolidated version consists of the third edition (2005) and its amendment 1 (2012). Therefore, there is no need to order the amendment in addition to this publication. This document includes the contents of the corrigenda 1 (2006) and 2 (2007), the contents of the corrigendum to Amendment 1 (2014), as well as the interpretation sheets 1 (2008), 2 (2009) and 3 (2013).

IEC 60601-1-2:2014
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 60601-1-2:2014 applies to the basic safety and essential performance of Medical Equipment (ME) equipment and ME systems in the presence of electromagnetic disturbances and to electromagnetic disturbances emitted by me equipment and me systems. This collateral standard to IEC 60601-1 specifies general requirements and tests for basic safety and essential performance with regard to electromagnetic disturbances and for electromagnetic emissions of ME equipment and ME systems. They are in addition to the requirements of the general standard IEC 60601-1 and serve as the basis for particular standards. This fourth edition cancels and replaces the third edition of IEC 60601-1-2, and constitutes a technical revision. The most significant changes with respect to the previous edition include the following modifications:
- specification of immunity test levels according to the environments of intended use, categorized according to locations that are harmonized with IEC 60601-1-11: the professional healthcare facility environment, the home healthcare environment and special environments;
- specification of tests and test levels to improve the safety of medical electrical equipment and medical electrical systems when portable RF communications equipment is used closer to the medical electrical equipment than was recommended based on the immunity test levels that were specified in the third edition;
- specification of immunity tests and immunity test levels according to the ports of the medical electrical equipment or medical electrical system;
- specification of immunity test levels based on the reasonably foreseeable maximum level of electromagnetic disturbances in the environments of intended use, resulting in some immunity test levels that are higher than in the previous edition; and
- better harmonization with the risk concepts of basic safety and essential performance, including deletion of the defined term "life-supporting". This new edition includes the following main additions:
- guidance for determination of immunity test levels for special environments;
- guidance for adjustment of immunity test levels when special considerations of mitigations or intended use are applicable;
- guidance on risk management for basic safety and essential performance with regard to electromagnetic disturbances; and
- guidance on identification of immunity pass/fail criteria.

IEC 60601-1-3:2008+AMD1:2013
Medical electrical equipment - Part 1-3: General requirements forbasic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

IEC 60601-1-3:2008+A1:2013 applies to X-ray equipment and to subassemblies of such equipment, where radiological images of a human patient are used for diagnosis, planning or guidance of medical procedures. The object of this collateral standard is to establish general requirements for protection against X-radiation in X-ray equipment, in order that the irradiation of the human patient, the operator, staff and members of the public can be kept as low as reasonably achievable, without jeopardizing the benefit of the radiological procedure. Particular standards may specify their appropriate values and/or measures for general requirements specified in this collateral standard. The implementation of the general requirements or the reference to the particular standard instead, shall be justified in the risk management process. This collateral standard considers radiation protection aspects related to X-radiation only. Requirements for the control of the electrical energy used to generate X-radiation, which is also an important aspect of radiation protection, are included in IEC 60601-1 and in particular standards for the safety and essential performance of the equipment concerned. This document cancels and replaces the first edition of IEC 60601-1-3, published in 1994 (which replaced IEC 60407 issued in 1973). It constitutes a technical revision. This edition has been restructured and aligned to IEC 60601-1(2005) and focussed on general requirements for RADIATION PROTECTION that apply to all diagnostic X-RAY EQUIPMENT. Requirements particular to specific equipment have been removed and will be covered in particular standards. For a description of the changes, see the mapping in Annex C. This consolidated version consists of the second edition (2008) and its amendment 1 (2013). Therefore, no need to order amendment in addition to this publication.

IEC 60601-1-6:2010+AMD1:2013+AMD2:2020
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

IEC 60601-1-6:2010+A1:2013+A2:2020 specifies a process for a manufacturer toanalyse, specify, design, verify and validate usability, as it relates to basic safety and essential performance of medical electrical equipment. This usability engineering process assesses and mitigates risks caused by usability problems associated with correct use and use errors, i.e., normal use. It can be used to identify but does not assess or mitigate risks associated with abnormal use. If the usability engineering process detailed in this collateral standard has been complied with and the acceptance criteria documented in the usability validation plan have been met (see 5.9 of IEC 62366:2007), then the residual risks, as defined in ISO 14971, associated with usability of me equipment are presumed to be acceptable, unless there is objective evidence to the contrary (see 4.1.2 of IEC 62366:2007). The object of this collateral standard is to specify general requirements that are in addition to those of the general standard and to serve as the basis for particular standards. This document cancels and replaces the second edition of IEC 60601-1-6 which has been technically revised. It was revised to align with the usability engineering process in IEC 62366. To allow for equipment manufacturers and testing organizations to make products and to equip themselves for conducting revised tests in accordance with this third edition, it is recommended by SC 62A that the content of this document not be adopted for mandatory implementation earlier than 3 years from the date of publication for equipment newly designed and not earlier than 5 years from the date of publication for equipment already in production. This consolidated version consists of the third edition (2010), its amendment 1 (2013) and its amendment 2 (2020). Therefore, no need to order amendments in addition to this publication

IEC 60601-1-8:2006+AMD1:2012+AMD2:2020
Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

IEC 60601-1-8:2006+A1:2012+A2:2020 Specifies basic safety and essential performance requirements and tests for alarm systems in medical electrical equipment and medical electrical systems and to provide guidance for their application. This is accomplished by defining alarm categories (priorities) by degree of urgency, consistent alarm signals and consistent control states and their marking for all alarm systems. This consolidated version consists of the second edition (2006), its amendment 1 (2012) and its amendment 2 (2020). Therefore, no need to order amendments in addition to this publication.

IEC 60601-1-9:2007+AMD1:2013+AMD2:2020
Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral Standard: Requirements for environmentally conscious design

IEC 60601-1-9:2007+A1:2013+A2:2020 The objective of this collateral standard is to improve the environmental impact for the entire range of medical electrical equipment, taking into account all stages of the product life cycle:
- product specification;
- design;
- manufacturing;
- sales, logistics, installation;
- use;
- end of life management. This means protecting the environment and human health from hazardous substances, conserving raw materials and energy, minimizing the generation of waste, as well as minimizing the adverse environmental impacts associated with waste. The criteria needed to reach this goal must be integrated into all stages of the medical electrical equipment life cycle from the specification stage to end of life management. The environmental impacts of me equipment through all life-cycle stages are determined from the medical electrical equipment's environmental aspects defined during the identification of need, product planning, and design stages. Consideration of environmental aspects as early as possible in these stages can produce numerous benefits that might include lower costs, stimulation of innovation and creativity, and increased knowledge about the product. It can also provide new business opportunities, and improved product quality as well as reduction of adverse environmental impacts. The assessment of the environmental aspects and impacts of medical electrical equipment is a developing science and it is anticipated that this collateral standard will require periodic updating as the science develops. The requirements given in this collateral standard do not replace national or international laws and regulations. Environmental protection is one element of the overall risk management process as required by the general standard. The acceptability of medical electrical equipment's environmental impacts are balanced against other factors, such as the product's intended function, performance, safety, cost, marketability, quality, legal and regulatory requirements. This balance can differ depending on the intended function of the medical electrical equipment. For example, a solution appropriate for life-saving or life-supporting medical electrical equipment might not be appropriate for a device intended to correct a minor ailment. A manufacturer of medical electrical equipment might have to justify, as a result of risk management, that a medical benefit outweighs the associated adverse environmental impacts. This consolidated version consists of the first edition (2007), its amendment 1 (2013) and its amendment 2 (2020). Therefore, no need to order amendments in addition to this publication.

IEC 60601-1-10:2007+AMD1:2013+AMD2:2020
Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers

IEC 60601-1-10:2007+A1:2013+A2:2020 Specifies requirements for the development (analysis, design, verification and validation) of a physiologic closed-loop controller (PCLC) as part of a physiologic closed-loop control system (PCLCS) in medical electrical equipment and medical electrical systems to control a physiologic variable. This collateral standard applies to various types of PCLC, e.g. linear and non-linear, adaptive, fuzzy, neural networks. This collateral standard applies to a closed-loop controller that sets the controller output variable in order to adjust (i.e., change or maintain) the measured physiologic variable by relating it to the reference variable. This consolidated version consists of the first edition (2007), its amendment 1 (2013) and its amendment 2 (2020). Therefore, no need to order amendments in addition to this publication.

IEC 60601-1-11:2015+AMD1:2020 
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-11:2015+A1:2020 applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems for use in the home healthcare environment. It applies regardless of whether the medical electrical equipment or medical electrical system is intended for use by a lay operator or by trained healthcare personnel. The home healthcare environment includes:
- the dwelling place in which a patient lives;
- other places where patients are present both indoors and outdoors, excluding professional healthcare facility environments where operators with medical training are continually available when patients are present.
This second edition cancels and replaces the first edition of IEC 60601-1-11, published in 2010, and constitutes a technical revision. The most significant changes with respect to the previous edition include the following modifications:
- correction of test method for relative humidity control at temperatures above 35 °C;
- redrafting of subclauses that altered instead of adding to the general standard or other collateral standards; and
- harmonizing with the changes to the amendments to the general standard and other collateral standards.
This consolidated version consists of the second edition (2015) and its amendment 1 (2020). Therefore, no need to order amendment in addition to this publication.

IEC 60601-1-12:2014+AMD1:2020 
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment

IEC 60601-1-12:2014+A1:2020 constitutes a collateral standard to IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance hereafter referred to as the general standard. Medical practice is increasingly using medical electrical equipment and medical electrical systems for monitoring, treatment or diagnosis of patients in the emergency medical services environment. The safety of medical electrical equipment in this uncontrolled, rough environment is a cause for concern. This collateral standard was developed with contributions from clinicians, engineers and regulators. The terminology, requirements, general recommendations and guidance of this collateral standard are intended to be useful for manufacturers of medical electrical equipment and medical electrical systems and for technical committees responsible for the development of particular standards. This International Standard applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems, hereafter referred to as ME equipment and ME systems, which are intended, as indicated in the instructions for use by their manufacturer, for use in the EMS environment (Emergency Medical Services environment). The object of this collateral standard is to provide general requirements for ME equipment and ME systems carried to the scene of an emergency and used there, as well as in transport, in situations where the ambient conditions differ from indoor conditions. The object of this collateral standard is to specify general requirements that are in addition to those of the general standard and to serve as the basis for particular standards. This consolidated version consists of the first edition (2014) and its amendment 1 (2020). Therefore, no need to order amendment in addition to this publication.

IEC TR 62348:2012 
Assessment of the impact of the most significant changes in Amendment 1 to IEC 60601-1:2005 and mapping of the clauses of IEC 60601-1:2005 to the previous edition

IEC/TR 62348:2012(E) which is a technical report, provides a tool to assist users of IEC 60601-1:2005 to assess the impact of the most significant changes in Amendment 1:2012. This technical report also provides a tool to assist users of IEC 60601-1 to trace requirements between the third edition and their source in the documents that form the basis of the third edition; principally the second edition as amended. This report is intended to be used by:
- those who must align standards based on the second edition of IEC 60601-1 with the third edition as amended;
- manufacturers of medical electrical equipment or medical electrical systems; and
- health care regulatory authorities, test houses and other organizations responsible for implementing standards for medical electrical equipment and medical electrical systems. This second edition cancels and replaces the first edition published in 2006. The second edition retains the mapping that traces the requirements of IEC 60601-1:2005 and its Amendment A1:2012 (Edition 3.1) from their source in the documents that relate to IEC 60601-1:1998 and its amendments (Edition 2.2). The second edition adds an assessment of the impact of the most significant changes in Amendment 1:2012.
...
segue in allegato

Certifico Srl - IT | Rev. 00.2020
©Copia autorizzata Abbonati

Collegati

Tags: Marcatura CE Abbonati Marcatura CE Regolamento Dispositivi Medici

Ultimi archiviati Marcatura CE

Ott 29, 2024 117

Decisione (UE) 2024/2779

Decisione (UE) 2024/2779 ID 22821 | 29.10.2024 Decisione (UE) 2024/2779 della Commissione, del 24 ottobre 2024, che istituisce il gruppo di esperti sulla progettazione ecocompatibile dei prodotti sostenibili e sull’etichettatura energetica (forum sulla progettazione ecocompatibile) (GU L 2024/2779… Leggi tutto
Ott 22, 2024 219

Regolamento di esecuzione (UE) 2024/994

Regolamento di esecuzione (UE) 2024/994 / Ecodesign banca dati dei prodotti ID 22773 | 22.10.2024 Regolamento di esecuzione (UE) 2024/994 della Commissione del 2 aprile 2024 recante i dettagli operativi della banca dati dei prodotti istituita a norma del regolamento (UE) 2017/1369 del Parlamento… Leggi tutto
Set 21, 2024 2486

Decreto 17 settembre 2024 n. 133

Decreto 17 settembre 2024 n. 133 ID 22594 | 21.09.2024 Decreto 17 settembre 2024 n. 133 - Regolamento di modifica al decreto 29 luglio 2008, n. 146 concernente il regolamento di attuazione dell'articolo 65 del decreto legislativo 18 luglio 2005, n. 171, recante il codice della nautica da diporto.… Leggi tutto

Più letti Marcatura CE

Nov 08, 2024 117412

Regolamento Prodotti da Costruzione (UE) 305/2011 - CPR

Regolamento CPR (UE) 305/2011 Regolamento (UE) N. 305/2011 del Parlamento Europeo e del Consiglio del 9 marzo 2011 che fissa condizioni armonizzate per la commercializzazione dei prodotti da costruzione e che abroga la direttiva 89/106/CEE del Consiglio (Testo rilevante ai fini del SEE) (GU L 88/10… Leggi tutto